欧美影视一区二区三区-欧美永久免费-欧美呦呦-欧美呦呦在线-午夜免费视频网站-午夜免费小视频

產品分類

Product category

首頁 > 技術支持 > DRX-I-PB

DRX-I-PB

點擊次數:3064     更新時間:2015-07-02
  1. Summary

The thermal conductivity is a significant parameter for researching physical properties of a material. In aviation area, atomic energy area, and industry department of building material or non-metallic material, an actual measurement or a forecast is required for the related material. There are steady testing method and dynamic testing method. The equipment adopts modified guarded hot plate method which is in line with the national standard. The whole testing process is finished either automatic or manual which satisfied the requirement of a high precision test. Reference standard: GB/T 10294-2008.

 

  1. Technical features

Applications: The equipment applies to test the thermal conductivity of clintheriform homogeneous material under dry or different moisture conditions. Thermal conductivity range: 0.015-2 W/M*K

  1. Can complete the test of plate method. The experiment temperature of equipment can be controlled. Temperature range: room temperature - 80.
  2. Digitizing: Instrument temperature degree better than 0.2, room temperature compensation automatic electronic. May also adopt external freezing temperature compensation.
  3. Power supply220V/50HZ; Adopt high precision regulated power supply.
  4. The result accuracy  ±3%

repeatability  ±2%

  1. Metering heating power is adjustable, constant current output 0-3A150v; accuracy: ±1%.
  2. Have connected PC, can accomplish computer automatic testing. Under the steady state conditions, it only 3 minutes to test one set of data.
  3. Can display the experiment parameters and curves.
  4. Can be done through a variety of test methods; make a reference and help to further comparative study on the properties of material thermal conductivity.
  5. The working conditions:

Environment temperature: 10°35

Relative humidity: ≤80%RH

③Require to operate in the condition of a stable room temperature. The average daily temperature fluctuations ≤±1.5

  1. Sample size and the effective area requirements  300×300×(10-40) mm
  2. Required the surface between the sample and the plate is flat and level. (Which can be connected by thermal conductive paste)

 

  1. Instrument composition

Instrument consist of the built-in high precision regulated power supply, thermometer, hot plate heater, the main protection of hot plate, temperature measuring thermocouple, cold plate constant temperature water bath, computer testing system.

a. Specimen part: including constant temperature device, through room temperature to 80

b. Heating system: heater of guarded hot plate temperature and heater of measuring power.

c. Temperature measurement system: adopting the high precision digital display instrument to ensure the accuracy and stability, and realize zero internal compensation.

d. measuring power adopting constant heating and digital display, the accuracy is better than 1%.

e. computer test parts including touch screen, communications components, and software.

f. instrument panels and the various parts see figure 1:

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 


Figure 1

1Power switch     2Work stop     3Work start     4Power light    5Power regulation

6Touch screen    7Samples room     8Cold plate lifting knob      9Base seat   

10Rolling wheel

 

 

 

  1. Operating steps
  1. Specimen preparation

Preparing samples according to the following requirements:

Sampling should be homogeneous parts from the sample material.

Sampling size is not greater than 300 x 300 x (10 ~ 40) mm

Smooth the sample’s surface after sampling.

Two surfaces of the specimen should be parallel, and uniformity of thickness. When testingthe surface of sample should be in contact with the plate flat and tight. If contact is not good, can coat the sample surface with a layer of powdery materials of the same material or high temperature heat conduction glue (cannot contain impurities and dust).

Before the test, sample thickness must be measured with a vernier caliper and accurate to 0.01 mm. Or use other ways to measure the sample thickness, detect four points on the sample and calculate for average thickness (unit mm), and record it for use during the test.

When the sample is foam or cotton of easily compressed material, please prepare four cylindrical solid pillars according to the thickness of the sample, used for testing.

  1. Sample test method
    • Sample loading:

Turning cold plate lifting handle and lower cold plate down, then place the sample on the center of the cold plate. If the sample is foam or cotton (easily compressed material), placing the four cylindrical solid pillars which has be prepared before in the corners of the sample, in order to prevent compression tight sample when rising cold plate.

Set the thermostatic water bath temperature

a

c

b

d

f

e

a

(figure2)

 

Open the power switch, cycle switch,, heat switch, cooling switch of the thermostatic water bath  (see figure 2).

 

apower 

bcycle

ccooling

dset

eup

fdown

Press “set" button on the thermostatic water bath plate, enter temperature setting mode (digital section indicates “PV", “SV" section displays the setting temperature), now press “up" or “down" to change the setting temperature. After done with setting, press “set" to finish (usually set as 20C).

  1. experiment mode

Rotate the power knob on the equipment to power on. Now the touch screen is on and enters the start interface. Click “enter" to enter the experiment interface. At the same time, press “start" button on the equipment. Touch the “start experiment" button on the screen. Wait until the temperature display section shows data, click and set the experimental temperature (usually 40?C) for the hot plate. Fill out the thickness of the sample, system will enter the experiment mode.

Guarded hot plate temperature control           

 

 

 

 

 

 

 

 

 

 


  1. Adjust voltage

When equipment starts “experiment mode", the temperatures of the major hot plate, guarded hot plate, and cold plate will display on the surface. Wait and observe the temperature change. When the temperature is stable (change amount ≤±0.3?C), observe the temperature differences between the major hot plate and the guarded hot plate (usually ≥0.2?C). Rotate “adjust voltage" knob clockwise. Observe the data displaying on the surface. Stop rotating when it shows about 3V and observe the temperature change. When the temperature is stable, observe if the displaying data close to the temperature of guarded hot plate (error less than ±0.2?C). If is, adjust voltage is accomplished. If not, rotate “adjust voltage" again and add more appropriate amount of voltage (adjustment according to the temperature difference between major and guarded hot plate, if the difference is large, adjust 3-5V more; if small, adjust 1-3V more) judge whether the two temperature are balance after the data is stable. When they are overheated, reduce the voltage immediay. Adjustment of voltage does not have to be integer.

  1. testing

When the temperatures of major and guarded hot plates are balanced, system would prompt a dialog box suggesting the temperatures are balanced. Click “ok" and click “start testing" to enter the testing mode. Wait about 3 min, the system will calculate one data. When obtained 4 data, the system will automatically calculate an average value.

When the average value comes out, users should wait 10-30min to click the “finish testing" and finish the test. After done with the testing, users could back to the initial surface. Rotate the “adjust voltage" to minimum (zero), press “stop" button on equipment, and turn off the power.

  1. tips
  1. When the thermostatic water bath sink is not filled with medium, do not turn it so it will not burn down the heating component.
  2. Interior parameters are set correctly before leaving factory, do not adjust freely.
  3. Turn off the power for all components, rotate the “adjust voltage" knob to zero, take the sample out, and clean the place where sample was. Do not scrub with hard objects.
  4. Check the connection of all lines once in a while. If the system is testing but all temperature data is zero, it might be because the connection is failed.
  1. data process and output
  1.  The testing system bases on the guarded hot plate apparatus and made some improvements on the mathematic model; adopts PC to test with the mathematic model.
  2. The method is on the basis of the measurements of heat transfer for two sides of the sample under the stable temperature condition to calculate the thermal conductivity coefficient.
  3.  Formula

== =  (unit: W/m.k)

I: heating current (A)

U: heating voltage (V)

d: sample thickness (m)

S: sample valid heating area (m2)

?Z: measure time interval (h)

?T: average temperature difference between two sides of the sample (? C)

Results should be accurate to three significant digits after the decimal point.

  1. The test requires accurate sample preparation to make sure the accuracy of the final result.
  2. When connecting external power source, ensure the reliability and grounding performance is good.

 

  1. Product list
  1. Testing host machine                        1
  2. High accurate thermostatic water bath           1
  3. Testing software (built-in in system)            1
  4. Operating manual                           1
  5. Certification of quality                       1         
在線咨詢 在線咨詢 聯系方式 二維碼

服務熱線

13607325732

掃一掃,關注我們

主站蜘蛛池模板: 久久精品店| 久久大香萑太香蕉综合网| 国产精品午夜久久| 女人l8毛片a一级毛片免费| 国产嫩草影院在线观看| 国产香蕉成人综合精品视频| 韩国一级毛片a级免观看| 国产区一区二区三| 欧美3d人妖交| 国语精品91自产拍在线观看二区| 国产精品国产国产aⅴ| 久久99久久精品免费思思| 福利精品在线| 欧美seav在线| 国产三级视频在线| 国产精品国产三级国产普通话对白 | 免费av一区二区三区| 久久五月网| 久久蜜桃视频| 成人久久精品| 久久精品国产只有精品6| 久久成年片色大黄全免费网站| 美女被啪到哭网站在线观看| 国产婷婷综合丁香亚洲欧洲| 国产手机在线观看视频| 国产一区亚洲| 成人久久久精品乱码一区二区三区| 国产偷v国产偷v亚洲高清| 久久久精品一区| 免费一级网站| 久久a 热6| 国产h在线播放| 麻豆国产在线观看一区二区| 韩国一大片a毛片| 国产hs免费高清在线观看| 久久道| 美女性生活毛片| 国产在线a| 国产精品福利在线观看| 黄网站在线播放| 毛片一区|